Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | G6NI1211AA | G6NI1211AB |
ReagentKit | |
Duration of the Analyzing | Month (2016/03) | CUM (2013/03/01~2019/09/30) | Month (2016/03) | CUM (2013/03/01~2019/09/30) |
Runs (N) | 30 | 760 | 30 | 760 |
Mean (U/gHb) | 2.1 | 2.1 | 6.2 | 5.8 |
SD | 0.15 | 0.18 | 0.24 | 0.51 |
CV (%) | 7.1 | 8.6 | 3.9 | 8.8 |
Target Value (U/gHb) | 1.3 | 1.3 | 4.9 | 4.9 |
Total Error (%) | 75.8 | 78.7 | 34.3 | 36.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | -5.9 | -4.8 | -1.7 | 0.2 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 1.3 ; SD : 0.09 ( ●) | |
Lot No.: G6NI1211AB ; Duration : 2013-03-01 to 2019-09-30 ; Target : 4.9 ; SD : 0.34 ( □) | |
|
Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 1.3 ; SD : 0.09 | |
|
Lot No.: G6NI1211AB ; Duration : 2013-03-01 to 2019-09-30 ; Target : 4.9 ; SD : 0.34 | |
|