Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | G6NI1703AI | G6NI1703AJ |
ReagentKit | |
Duration of the Analyzing | Month (2020/01) | CUM (2017/05/01~2020/01/31) | Month (2020/01) | CUM (2017/05/01~2020/01/31) |
Runs (N) | 0 | 191 | 0 | 191 |
Mean (U/gHb) | - | 1.8 | - | 5.9 |
SD | - | 0.26 | - | 0.64 |
CV (%) | - | 14.4 | - | 10.8 |
Target Value (U/gHb) | 1.7 | 1.7 | 5.4 | 5.4 |
Total Error (%) | - | 34.8 | - | 31.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 1.0 | - | 1.0 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2020-01-31 ; Target : 1.7 ; SD : 0.17 ( ●) | |
Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2020-01-31 ; Target : 5.4 ; SD : 0.54 ( □) | |
|
Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2020-01-31 ; Target : 1.7 ; SD : 0.17 | |
|
Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2020-01-31 ; Target : 5.4 ; SD : 0.54 | |
|