Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | G6NI2303BA | G6NI2303BB |
ReagentKit | |
Duration of the Analyzing | Month (2025/07) | CUM (2024/02/01~2025/07/31) | Month (2025/07) | CUM (2024/02/01~2025/07/31) |
Runs (N) | 99 | 400 | 99 | 400 |
Mean (U/gHb) | 1.3 | 1.4 | 4.1 | 4.2 |
SD | 0.09 | 0.13 | 0.24 | 0.40 |
CV (%) | 6.9 | 9.3 | 5.9 | 9.5 |
Target Value (U/gHb) | 1.4 | 1.4 | 4.0 | 4.0 |
Total Error (%) | 21.0 | 18.6 | 14.2 | 24.0 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 1.9 | 2.2 | 3.0 | 1.6 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: G6NI2303BA ; Duration : 2024-02-01 to 2025-07-31 ; Target : 1.4 ; SD : 0.13 ( ●) | |
Lot No.: G6NI2303BB ; Duration : 2024-02-01 to 2025-07-31 ; Target : 4.0 ; SD : 0.39 ( □) | |
 |
Lot No.: G6NI2303BA ; Duration : 2024-02-01 to 2025-07-31 ; Target : 1.4 ; SD : 0.13 | |
 |
Lot No.: G6NI2303BB ; Duration : 2024-02-01 to 2025-07-31 ; Target : 4.0 ; SD : 0.39 | |
 |
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated