Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| | CI | CII |
| QC Control Lot No. | G6NI2401BC | G6NI2401BD |
| ReagentKit | PE (ND-1000) |
| Duration of the Analyzing | Month (2025/11) | CUM (2025/02/01~2025/11/30) | Month (2025/11) | CUM (2025/02/01~2025/11/30) |
| Runs (N) | 175 | 725 | 175 | 725 |
| Mean (U/gHb) | 11.3 | 4.0 | 4.9 | 5.0 |
| SD | 127.70 | 62.74 | 0.47 | 0.35 |
| CV (%) | 1130.1 | 1568.5 | 9.6 | 7.0 |
| Target Value (U/gHb) | 1.7 | 1.7 | 5.0 | 5.0 |
| Total Error (%) | 2824.9 | 3272.3 | 21.2 | 14.0 |
| TEa (%) | 20 | 20 | 20 | 20 |
| σ | -0.5 | -0.1 | 1.9 | 2.9 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
| Lot No.: G6NI2401BC ; Duration : 2025-02-01 to 2025-11-30 ; Target : 1.7 ; SD : 0.14 ( ●) | |
| Lot No.: G6NI2401BD ; Duration : 2025-02-01 to 2025-11-30 ; Target : 5.0 ; SD : 0.31 ( □) | |
 |
| Lot No.: G6NI2401BC ; Duration : 2025-02-01 to 2025-11-30 ; Target : 1.7 ; SD : 0.14 | |
 |
| Lot No.: G6NI2401BD ; Duration : 2025-02-01 to 2025-11-30 ; Target : 5.0 ; SD : 0.31 | |
 |