| | CI | CII |
| QC Control Lot No. | G6NI2401BC | G6NI2401BD |
| ReagentKit | PE (ND-1000) |
| Duration of the Analyzing | Month (2025/10) | CUM (2025/02/01~2025/10/31) | Month (2025/10) | CUM (2025/02/01~2025/10/31) |
| Runs (N) | 42 | 119 | 42 | 119 |
| Mean (U/gHb) | 1.7 | 1.8 | 5.2 | 5.1 |
| SD | 0.16 | 0.14 | 0.36 | 0.38 |
| CV (%) | 9.4 | 7.8 | 6.9 | 7.5 |
| Target Value (U/gHb) | 1.8 | 1.8 | 5.2 | 5.2 |
| Total Error (%) | 24.4 | 15.6 | 13.8 | 16.8 |
| TEa (%) | 20 | 20 | 20 | 20 |
| σ | 1.5 | 2.6 | 2.9 | 2.4 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
| Lot No.: G6NI2401BC ; Duration : 2025-02-01 to 2025-10-31 ; Target : 1.8 ; SD : 0.09 ( ●) | |
| Lot No.: G6NI2401BD ; Duration : 2025-02-01 to 2025-10-31 ; Target : 5.2 ; SD : 0.38 ( □) | |
 |
| Lot No.: G6NI2401BC ; Duration : 2025-02-01 to 2025-10-31 ; Target : 1.8 ; SD : 0.09 | |
 |
| Lot No.: G6NI2401BD ; Duration : 2025-02-01 to 2025-10-31 ; Target : 5.2 ; SD : 0.38 | |
 |