Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | G6NI1211AA | G6NI1211AB |
ReagentKit | |
Duration of the Analyzing | Month (2014/03) | CUM (2013/03/01~2019/09/30) | Month (2014/03) | CUM (2013/03/01~2019/09/30) |
Runs (N) | 240 | 3565 | 240 | 3565 |
Mean (U/gHb) | 1.9 | 2.0 | 5.7 | 5.7 |
SD | 0.12 | 0.19 | 0.37 | 0.45 |
CV (%) | 6.3 | 9.5 | 6.5 | 7.9 |
Target Value (U/gHb) | 2.0 | 2.0 | 6.0 | 6.0 |
Total Error (%) | 17.6 | 19.0 | 18.0 | 20.8 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 2.4 | 2.1 | 2.3 | 1.9 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 2.0 ; SD : 0.10 ( ●) | |
Lot No.: G6NI1211AB ; Duration : 2013-03-01 to 2019-09-30 ; Target : 6.0 ; SD : 0.16 ( □) | |
|
Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 2.0 ; SD : 0.10 | |
|