Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | G6NI1211AA | G6NI1211AB |
ReagentKit | |
Duration of the Analyzing | Month (2018/04) | CUM (2013/03/01~2019/09/30) | Month (2018/04) | CUM (2013/03/01~2019/09/30) |
Runs (N) | 21 | 984 | 21 | 984 |
Mean (U/gHb) | 2.1 | 2.2 | 6.2 | 6.2 |
SD | 0.15 | 0.19 | 0.24 | 0.31 |
CV (%) | 7.1 | 8.6 | 3.9 | 5.0 |
Target Value (U/gHb) | 2.1 | 2.1 | 6.4 | 6.4 |
Total Error (%) | 14.3 | 22.0 | 10.9 | 13.1 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 2.8 | 1.8 | 4.3 | 3.4 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 2.1 ; SD : 0.2 ( ●) | |
Lot No.: G6NI1211AB ; Duration : 2013-03-01 to 2019-09-30 ; Target : 6.4 ; SD : 0.6 ( □) | |
|
Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 2.1 ; SD : 0.2 | |
|
Lot No.: G6NI1211AB ; Duration : 2013-03-01 to 2019-09-30 ; Target : 6.4 ; SD : 0.6 | |
|