Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | G6NI2101AQ | G6NI2101AR |
ReagentKit | PE (ND-1000) |
Duration of the Analyzing | Month (2022/07) | CUM (2021/09/01~2024/09/30) | Month (2022/07) | CUM (2021/09/01~2024/09/30) |
Runs (N) | 0 | 900 | 0 | 900 |
Mean (U/gHb) | - | 1.8 | - | 5.1 |
SD | - | 0.28 | - | 0.35 |
CV (%) | - | 15.6 | - | 6.9 |
Target Value (U/gHb) | 1.7 | 1.7 | 4.9 | 4.9 |
Total Error (%) | - | 37.0 | - | 17.8 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | - | 0.9 | - | 2.3 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: G6NI2101AQ ; Duration : 2021-09-01 to 2024-09-30 ; Target : 1.7 ; SD : 0.10 ( ●) | |
Lot No.: G6NI2101AR ; Duration : 2021-09-01 to 2024-09-30 ; Target : 4.9 ; SD : 0.30 ( □) | |
|
Lot No.: G6NI2101AQ ; Duration : 2021-09-01 to 2024-09-30 ; Target : 1.7 ; SD : 0.10 | |
|
Lot No.: G6NI2101AR ; Duration : 2021-09-01 to 2024-09-30 ; Target : 4.9 ; SD : 0.30 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020