Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | G6NI2101AQ | G6NI2101AR |
ReagentKit | PE (ND-1000) |
Duration of the Analyzing | Month (2022/10) | CUM (2021/09/01~2024/09/30) | Month (2022/10) | CUM (2021/09/01~2024/09/30) |
Runs (N) | 173 | 3273 | 173 | 3273 |
Mean (U/gHb) | 1.5 | 1.4 | 4.9 | 4.4 |
SD | 0.09 | 0.13 | 1.12 | 0.39 |
CV (%) | 6.0 | 9.3 | 22.9 | 8.9 |
Target Value (U/gHb) | 1.5 | 1.5 | 4.5 | 4.5 |
Total Error (%) | 12.0 | 25.2 | 54.6 | 19.9 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 3.3 | 1.4 | 0.5 | 2.0 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: G6NI2101AQ ; Duration : 2021-09-01 to 2024-09-30 ; Target : 1.5 ; SD : 0.15 ( ●) | |
Lot No.: G6NI2101AR ; Duration : 2021-09-01 to 2024-09-30 ; Target : 4.5 ; SD : 0.45 ( □) | |
|
Lot No.: G6NI2101AQ ; Duration : 2021-09-01 to 2024-09-30 ; Target : 1.5 ; SD : 0.15 | |
|
Lot No.: G6NI2101AR ; Duration : 2021-09-01 to 2024-09-30 ; Target : 4.5 ; SD : 0.45 | |
|
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020