Internal Quality Control Program (IQC) for Neonatal G6PD Screening Test
G6PD IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| | CI | CII |
| QC Control Lot No. | G6NI1211AA | G6NI1211AB |
| ReagentKit | |
| Duration of the Analyzing | Month (2019/09) | CUM (2013/03/01~2019/09/30) | Month (2019/09) | CUM (2013/03/01~2019/09/30) |
| Runs (N) | 0 | 947 | 0 | 947 |
| Mean (U/gHb) | - | 2.0 | - | 5.7 |
| SD | - | 0.19 | - | 0.41 |
| CV (%) | - | 9.5 | - | 7.2 |
| Target Value (U/gHb) | 2.0 | 2.0 | 5.6 | 5.6 |
| Total Error (%) | - | 19.0 | - | 16.2 |
| TEa (%) | 20 | 20 | 20 | 20 |
| σ | - | 2.1 | - | 2.5 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
| Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 2.0 ; SD : 0.20 ( ●) | |
| Lot No.: G6NI1211AB ; Duration : 2013-03-01 to 2019-09-30 ; Target : 5.6 ; SD : 0.30 ( □) | |
 |
| Lot No.: G6NI1211AA ; Duration : 2013-03-01 to 2019-09-30 ; Target : 2.0 ; SD : 0.20 | |
 |
| Lot No.: G6NI1211AB ; Duration : 2013-03-01 to 2019-09-30 ; Target : 5.6 ; SD : 0.30 | |
 |
Contact us
Address:
P.O. Box 624 Taipei Xinwei, Taiwan, 106926 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 09 Apr, 2020