QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

Select LotNo:
NS013


 CICII
QC Control Lot No.G6NI1703AIG6NI1703AJ
ReagentKit
Duration of the AnalyzingMonth
(2018/09)
CUM
(2017/05/01~2019/06/30)
Month
(2018/09)
CUM
(2017/05/01~2019/06/30)
Runs (N)26247982624798
Mean (U/gHb)1.71.75.55.5
SD0.080.130.250.27
CV (%)4.77.64.54.9
Target Value (U/gHb)1.71.75.55.5
Total Error (%)9.415.39.19.8
TEa (%)20202020
σ4.32.64.44.1
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)

Month :    ;    Cumulative : from to
[TOP]



CI and CII SDI
QC Chart

Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2019-06-30 ; Target : 1.7 ; SD : 0.10      ( ●) 
Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2019-06-30 ; Target : 5.5 ; SD : 0.30      ( □)
NS013
SDI Diagram of CI & CII

Month :    ;    Cumulative : from to
[TOP]



CI SDI QC Chart

Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2019-06-30 ; Target : 1.7 ; SD : 0.10
NS013
SDI Diagram of CI

Month :    ;    Cumulative : from to
[TOP]



CII SDI QC Chart

Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2019-06-30 ; Target : 5.5 ; SD : 0.30
NS013
SDI Diagram of CII

Month :    ;    Cumulative : from to
[TOP]