QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

Select LotNo:
NS015


 CICII
QC Control Lot No.G6NI2101AQG6NI2101AR
ReagentKitPE (ND-1000)
Duration of the AnalyzingMonth
(2023/02)
CUM
(2021/09/01~2024/06/30)
Month
(2023/02)
CUM
(2021/09/01~2024/06/30)
Runs (N)0327303273
Mean (U/gHb)-1.4-4.4
SD-0.13-0.39
CV (%)-9.3-8.9
Target Value (U/gHb)1.51.54.54.5
Total Error (%)-25.2-19.9
TEa (%)20202020
σ-1.4-2.0
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)

Month :    ;    Cumulative : from to
[TOP]



CI and CII SDI
QC Chart

Lot No.: G6NI2101AQ ; Duration : 2021-09-01 to 2024-06-30 ; Target : 1.5 ; SD : 0.15      ( ●) 
Lot No.: G6NI2101AR ; Duration : 2021-09-01 to 2024-06-30 ; Target : 4.5 ; SD : 0.45      ( □)
NS015
SDI Diagram of CI & CII

Month :    ;    Cumulative : from to
[TOP]



CI SDI QC Chart

Lot No.: G6NI2101AQ ; Duration : 2021-09-01 to 2024-06-30 ; Target : 1.5 ; SD : 0.15
NS015
SDI Diagram of CI

Month :    ;    Cumulative : from to
[TOP]



CII SDI QC Chart

Lot No.: G6NI2101AR ; Duration : 2021-09-01 to 2024-06-30 ; Target : 4.5 ; SD : 0.45
NS015
SDI Diagram of CII

Month :    ;    Cumulative : from to