QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

Select LotNo:
NS019


 CICII
QC Control Lot No.G6NI1703AIG6NI1703AJ
ReagentKit
Duration of the AnalyzingMonth
(2018/09)
CUM
(2017/05/01~2019/06/30)
Month
(2018/09)
CUM
(2017/05/01~2019/06/30)
Runs (N)21924832192483
Mean (U/gHb)1.81.75.75.5
SD0.110.120.300.31
CV (%)6.17.15.35.6
Target Value (U/gHb)1.81.85.75.7
Total Error (%)12.219.710.514.8
TEa (%)20202020
σ3.32.03.82.9
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)

Month :    ;    Cumulative : from to
[TOP]



CI and CII SDI
QC Chart

Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2019-06-30 ; Target : 1.8 ; SD : 0.11      ( ●) 
Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2019-06-30 ; Target : 5.7 ; SD : 0.34      ( □)
NS019
SDI Diagram of CI & CII

Month :    ;    Cumulative : from to
[TOP]



CI SDI QC Chart

Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2019-06-30 ; Target : 1.8 ; SD : 0.11
NS019
SDI Diagram of CI

Month :    ;    Cumulative : from to
[TOP]



CII SDI QC Chart

Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2019-06-30 ; Target : 5.7 ; SD : 0.34
NS019
SDI Diagram of CII

Month :    ;    Cumulative : from to
[TOP]