QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

Select LotNo:
NS021


 CICII
QC Control Lot No.G6NI1801AKG6NI1801AL
ReagentKit
Duration of the AnalyzingMonth
(2023/02)
CUM
(2019/02/01~2024/08/31)
Month
(2023/02)
CUM
(2019/02/01~2024/08/31)
Runs (N)0143601436
Mean (U/gHb)-1.3-5.4
SD-0.09-0.25
CV (%)-6.9-4.6
Target Value (U/gHb)1.31.35.55.5
Total Error (%)-13.8-11.1
TEa (%)20202020
σ-2.9-4.0
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)

Month :    ;    Cumulative : from to
[TOP]



CI and CII SDI
QC Chart

Lot No.: G6NI1801AK ; Duration : 2019-02-01 to 2024-08-31 ; Target : 1.3 ; SD : 0.08      ( ●) 
Lot No.: G6NI1801AL ; Duration : 2019-02-01 to 2024-08-31 ; Target : 5.5 ; SD : 0.25      ( □)
NS021
SDI Diagram of CI & CII

Month :    ;    Cumulative : from to
[TOP]



CI SDI QC Chart

Lot No.: G6NI1801AK ; Duration : 2019-02-01 to 2024-08-31 ; Target : 1.3 ; SD : 0.08
NS021
SDI Diagram of CI

Month :    ;    Cumulative : from to
[TOP]



CII SDI QC Chart

Lot No.: G6NI1801AL ; Duration : 2019-02-01 to 2024-08-31 ; Target : 5.5 ; SD : 0.25
NS021
SDI Diagram of CII

Month :    ;    Cumulative : from to
[TOP]