QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

QC Chart of Internal Quality Control (IQC)for G6PD Quantitative Test

Select LotNo:
NS022


 CICII
QC Control Lot No.G6NI1703AIG6NI1703AJ
ReagentKit
Duration of the AnalyzingMonth
(2018/09)
CUM
(2017/05/01~2019/06/30)
Month
(2018/09)
CUM
(2017/05/01~2019/06/30)
Runs (N)1519115191
Mean (U/gHb)2.21.86.65.9
SD0.140.260.240.64
CV (%)6.414.43.610.8
Target Value (U/gHb)1.71.75.45.4
Total Error (%)42.134.829.531.0
TEa (%)20202020
σ-1.51.0-0.61.0
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)

Month :    ;    Cumulative : from to
[TOP]



CI and CII SDI
QC Chart

Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2019-06-30 ; Target : 1.7 ; SD : 0.17      ( ●) 
Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2019-06-30 ; Target : 5.4 ; SD : 0.54      ( □)
NS022
SDI Diagram of CI & CII

Month :    ;    Cumulative : from to
[TOP]



CI SDI QC Chart

Lot No.: G6NI1703AI ; Duration : 2017-05-01 to 2019-06-30 ; Target : 1.7 ; SD : 0.17
NS022
SDI Diagram of CI

Month :    ;    Cumulative : from to
[TOP]



CII SDI QC Chart

Lot No.: G6NI1703AJ ; Duration : 2017-05-01 to 2019-06-30 ; Target : 5.4 ; SD : 0.54
NS022
SDI Diagram of CII

Month :    ;    Cumulative : from to
[TOP]